Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1827-2025

HEMOCUE INC F/GLUCOSE 201 ANALYZER (25/VL 4VL/BX) MICROCUVETTE. Recall

HEMOCUE INC F/GLUCOSE 201 ANALYZER (25/VL 4VL/BX) MICROCUVETTE.

FDA recall Class II

Ongoing

FDA lists this Class II recall for HEMOCUE INC F/GLUCOSE 201 ANALYZER (25/VL 4VL/BX) MICROCUVETTE.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1827-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 24, 2025
FDA report date
Jun 4, 2025
Quantity in commerce
4 units
Distribution
See key details
FDA cause
Storage

What is being recalled?

HEMOCUE INC F/GLUCOSE 201 ANALYZER (25/VL 4VL/BX) MICROCUVETTE.

Affected model & UDI checker

Search the exact identifiers published for Z-1827-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
110706HEMOCUE INC F/GLUCOSE 201 ANALYZER (25/VL 4VL/BX) MICROCUVETTE.Check source noticeNot specified
Not specifiedHEMOCUE INC F/GLUCOSE 201 ANALYZER (25/VL 4VL/BX) MICROCUVETTE.Check source notice17311091107067
FDA code information

Catalog No. 110706; GTIN: 17311091107067; Order No. 46085136.

View the original FDA recall notice

Why was this product recalled?

transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result.

What actions did Mckesson Medical-Surgical Inc. Corporate Office take?

An URGENT PRODUCT RECALL notification dated 4/21/25 was mailed to consignees. The notification instructs consignees to immediately examine their inventory and quarantine all product identified as subject to this recall action. Consignees are asked to return the completed response form to MMSQARecalls@McKesson.com and to destroy any quarantined devices. Questions about this recall can be directed to 1-800-688-8840 from 8:00 am to 5:00 pm Monday through Friday. If product has been further distributed, consignees are asked to forward the provided recall notice to those customers.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology