CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system. Recall
CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.
FDA recall Class II
Ongoing
FDA lists this Class II recall for CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1828-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 6, 2026
- FDA report date
- Apr 29, 2026
- Quantity in commerce
- 78 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.
Affected model & UDI checker
Search the exact identifiers published for Z-1828-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| C7000 | CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system. | Check source notice | 10381780126232 |
FDA code information
Model: C7000; UDI: 10381780126232; Serial Numbers: CHA1902101IE, CHA2001301IE, CHA2001401IE, CHA2002301IE, CHA2202301IE, CHA2305101IE, CHA2305401IE, CHA2307001IE, CHA2402201IE, CHB1711001IE, CHB1723601IE, CHB2000801IE, CHB2100901IE, CHB2301401IE, CHB2500101IE, CHC1800601IE, CHC1902601IE, CHC1902701IE, CHC2001201IE, CHD1705801IE, CHD1707101IE, CHD1801001IE, CHD1801101IE, CHD1902501IE , CHE1707201IE, CHE1707301IE, CHE2203301IE, CHE2204001IE, CHE2301001IE , CHE2305001IE, CHE2502901IE, CHE2503101IE, CHE2503201IE, CHE2503401IE, CHE2503701IE, CHF1701401IE, CHF1702701IE, CHF2100701IE, CHF2303701IE, CHF2305401IE, CHG1700101IE, CHG1700801IE, CHG1902401IE, CHG2000201IE, CHG2000301IE, CHG2204101IE, CHG2402301IE, CHG2502801IE, CHG2506001IE, CHH1802901IE, CHH2402801IE, CHH2500601IE, CHJ2001101IE, CHJ2001301IE , CHJ2301201IE, CHJ2304601IE, CHK1700801IE, CHK1702301IE, CHK1703001IE, CHK2402601IE, CHK2403401IE, CHK2404101IE, CHL1701701IE, CHL1900201IE, CHL2102901IE , CHL2103001IE, CHL2301301IE, CHL2302001IE, CHL2302101IE, CHL2403301IE, CHM1901901IE, CHM2100101IE, CHM2101601IE, CHM2200201IE, CHM2200301IE, CHM2200501IE, CHM2300701IE, CHM2402101IE.
Why was this product recalled?
Software issue that renders the touch screen unresponsive.
What actions did Integra LifeSciences Corp. (NeuroSciences) take?
An URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice dated 3/6/26 was sent to consignees. The notice instructs consignees to restart systems when freeze to restore functionality of the touchscreen capabilities for 8 minutes; this action may be used multiple times. Integra's Service and Repair can provide remote support to disable and re-enable the write filter with Integra employees and/or Authorized service personnel. This will reset the console to provide approximately 28 days with no touchscreen issue. Integra Lifesciences Service Team will contact consignees within 5 to 10 business days to schedule correction of affected devices at no cost. Consignees with any questions can direct them to Customer Service by emailing custsvcnj@integralife.com or by phone at 1-800-654-2873.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1828-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology