ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System Recall
ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System
FDA recall Class II
Ongoing
FDA lists this Class II recall for ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1857-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 22, 2024
- FDA report date
- Jun 4, 2025
- Quantity in commerce
- 24 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Part Number: 25001040E. UDI-DI B27825001040E0. Lot Numbers: 1989873, 1989874
Why was this product recalled?
Potential challenge with intraoperative assembly of the external taper of a single midsection with the internal taper of a mating proximal femur.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Onkos Surgical, Inc. take?
Onkos Medical notified consignees via email on about 09/22/2024. Consignees were instructed to segregate and quarantine any product on hand immediately, complete and return the response form, and arrange for existing implants to be returned.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1857-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026