Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1859-2025

Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system Recall

Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system

FDA recall Class II

Ongoing

FDA lists this Class II recall for Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1859-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 31, 2025
FDA report date
Jun 4, 2025
Quantity in commerce
96 systems
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system

Affected model & UDI checker

Search the exact identifiers published for Z-1859-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
6003000804Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning systemCheck source noticeNot specified
6003000805Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning systemCheck source noticeNot specified
Not specifiedUltrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning systemCheck source notice00816634020013
Not specifiedUltrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning systemCheck source notice00816634020006
FDA code information

Triton 36 Model Number 6003000804 UDI-DI code: 00816634020013 Serial Numbers: 1601-XL36-249-1 1604-XL36-277-1 1808-XL36TR-596-1 1810-XL36TR-618-1 1905-XL36TR-707-1 1907-XL36TR-721-1 1911-XL36TR-800-1 2001-XL36TR-834-1 2004-XL36TR-859-1 2004-XL36TR-873-1 2104-XL36TR-947-1 2201-XL36TR-001-1 2203-XL36TR-017-1 2306-XL36TR-050-1 2306-XL36TR-051-1 2307-XL36TR-053-1 2312-XL36TR-120-1 2401-XL36TR-001-1 2402-XL36TR-017-1 2402-XL36TR-018-1 2402-XL36TR-027-1 2402-XL36TR-028-1 2402-XL36TR-029-1 2402-XL36TR-030-1 2405-XL36TR-048-1 2405-XL36TR-049-1 2405-XL36TR-050-1 2405-XL36TR-051-1 2405-XL36TR-060-1 2405-XL36TR-061-1 2405-XL36TR-062-1 2405-XL36TR-063-1 2406-XL36TR-064-1 2406-XL36TR-065-1 2406-XL36TR-072-1 2406-XL36TR-074-1 2406-XL36TR-075-1 2407-XL36TR-079-1 2407-XL36TR-083-1 2409-XL36TR-101-1 2409-XL36TR-103-1 2409-XL36TR-105-1 2409-XL36TR-107-1 2409-XL36TR-109-1 2409-XL36TR-110-1 2410-XL36TR-111-1 2410-XL36TR-113-1 2410-XL36TR-114-1 2411-XL36TR-118-1 2411-XL36TR-122-1 2412-XL36TR-135-1 XL36TR4381 XL36TR4961 XL36TR5111 2307-XL36TR-055-1 Triton 72 Model Number 6003000805 UDI-DI code: 00816634020006 Serial Numbers: 1511-XL72-230-1 1606-XL72-291-1 1706-XL72TR-401-1 1805-XL72TR-544-1 1805-XL72TR-546-1 1806-XL72TR-564-1 1806-XL72TR-565-1 1807-XL72TR-568-1 1809-XL72TR-611-1 1910-XL72TR-788-1 2002-XL72TR-844-1 2004-XL72TR-878-1 2004-XL72TR-879-1 2005-XL72TR-881-1 2006-XL72TR-890-1 2103-XL72TR-929-1 2103-XL72TR-930-1 2104-XL72TR-946-1 2112-XL72TR-1022-1 2202-XL72TR-013-1 2202-XL72TR-014-1 2202-XL72TR-015-1 2203-XL72TR-016-1 2207-XL72TR-045-1 2211-XL72TR-088-1 2304-XL72TR-028-1 2305-XL72TR-030-1 2305-XL72TR-034-1 2309-XL72TR-073-1 2310-XL72TR-099-1 2402-XL72TR-023-1 2402-XL72TR-024-1 2404-XL72TR-047-1 2405-XL72TR-052-1 2405-XL72TR-053-1 2406-XL72TR-073-1 2407-XL72TR-080-1 2407-XL72TR-084-1 2409-XL72TR-104-1 2411-XL72TR-120-1 XL72TR4401

View the original FDA recall notice

Why was this product recalled?

During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation.

What actions did Ultra Clean Systems, Inc. take?

On 03/31/2025, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via FedEx to customers informing them that as a result of an investigation, it has been revealed that a combination of chemical and electrical interactions that result in electrochemical degradation, compromising the structural integrity of the sensor over time. This degradation could potentially lead to water leakage into the machine. On 04/25/2025, the firm sent via FedEx and updated letter. Customers are notified and/or instructed that/to: -A field technicians will be contacting you to arrange for replacement if your unit has been impacted by this issue. -Complete and sign the attached response form to acknowledge that they have received the notification. Return the completed form to Getinge by emailing a scanned copy to recallresponses.qrc@getinge.com Public contact for questions/assistance - recalls.qrc@getinge.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology