NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. Recall
NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.
FDA recall Class II
Ongoing
FDA lists this Class II recall for NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1864-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 28, 2025
- FDA report date
- Jun 4, 2025
- Quantity in commerce
- 115,465 cases (230,930 units)
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.
Affected model & UDI checker
Search the exact identifiers published for Z-1864-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| RFP-400 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. | Check source notice | Not specified |
| RFP-401 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. | Check source notice | Not specified |
| RFP-402 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. | Check source notice | Not specified |
| RFP-403 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. | Check source notice | Not specified |
| RFP-404 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. | Check source notice | Not specified |
| RFP-406 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. | Check source notice | Not specified |
| RFP-453 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. | Check source notice | Not specified |
| RFP-454 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. | Check source notice | Not specified |
| RFP-456 | NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456. | Check source notice | Not specified |
FDA code information
a. UDI-DI number: M535RFP4000. Device Catalog Number RFP-400. Lot Numbers: Q2407307, Q2407308, Q2407309. b. UDI-DI number: M535RFP4010. Device Catalog Number RFP-401. Lot Numbers: Q2407061, Q2407147, Q2407148, Q2407149, Q2407313, Q2407386, Q2407387, Q2407388, Q2407389, Q2407390. c. UDI-DI number: M535RFP4020. Device Catalog Number RFP-402. Lot Numbers: Q2407213, Q2407214. d. UDI-DI number: M535RFP4030. Device Catalog Number RFP-403. Lot Number: Q2407311. e. UDI-DI number: M535RFP4040. Device Catalog Number RFP-404. Lot Numbers: Q2407062, Q2407063, Q2407064, Q2407065, Q2407067, Q2407068, Q2407069, Q2407070, Q2407071, Q2407217, Q2407218, Q2407219, Q2407220, Q2407221, Q2407222, Q2407223, Q2407224, Q2407225, Q2407301, Q2407302, Q2407303, Q2407304, Q2407305, Q2407306, Q2407391, Q2407392, Q2407405, Q2407406. f. UDI-DI number: M535RFP4060. Device Catalog Number RFP-406. Lot Numbers: Q2407072, Q2407137, Q2407140, Q2407141, Q2407216, Q2407310, Q2407312. g. UDI-DI number: M535RFP4530. Device Catalog Number RFP-453. Lot Number: Q2407066. h. UDI-DI number: M535RFP4540. Device Catalog Number RFP-454. Lot Numbers: Q2407145, Q2407146, Q2407226. j. UDI-DI number: M535RFP4560. Device Catalog Number RFP-456. Lot Numbers: Q2407138, Q2407139, Q2407142, Q2407143, Q2407144, Q2407227.
Why was this product recalled?
Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior to treatment.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did NxStage Medical Inc take?
NxStage notified consignees via mail on 04/28/2025. Consignees were instructed to check inventory for affected units and cease use if any are on hand, complete and return the response form, and contact NxStage Customer Service to arrange return of all affected units.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1864-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology