VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides. Recall
VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides.
FDA recall Class II
Ongoing
FDA lists this Class II recall for VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1876-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 14, 2025
- FDA report date
- Jun 4, 2025
- Quantity in commerce
- 3,587 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides.
Affected model & UDI checker
Search the exact identifiers published for Z-1876-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 8097990 | VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides. | Check source notice | Not specified |
FDA code information
Catalog number: 8097990; GTIN#: 10758750004355; Coating Number: 1291; Lot No. 3788-1291-0019, 3788-1291-0052, 3788-1291-0078, 3788-1291-0079, 3788-1291-0080, 3788-1291-9975.
Why was this product recalled?
Product was incorrectly assigned GEN 88 instead of GEN 83 on product carton, foil wrapper, and cartridge (cart) label.
What actions did ORTHO-CLINICAL DIAGNOSTICS, INC. take?
An URGENT PRODUCT CORRECTION NOTIFICATION was mailed to consignees. The notification to distributors instructs consignees to forward the provided notification and Customer Confirmation Receipt to all customers who were shipped affected lots and to discontinue distributing affected lots. Affected lots in possession of distributors should be discarded. Distributors are asked to return a completed Distributor Confirmation Receipt form. The notification to customers instructs consignees to check their inventory and destroy any identified product subject to this recall. Consignees are instructed to post the provided notification near all VITROS devices and within device User Documentation until the issue has been resolved. A copy of the recall notification should be forwarded if a unit was distributed outside your facility. Consignees are to complete and return the provided Confirmation of Receipt form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1876-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026