Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262 Recall
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262
FDA recall Class II
Ongoing
FDA lists this Class II recall for Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1880-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 27, 2026
- FDA report date
- Apr 29, 2026
- Quantity in commerce
- 270,311 total
- Distribution
- See key details
- FDA cause
- Component change control
What is being recalled?
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262
Affected model & UDI checker
Search the exact identifiers published for Z-1880-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262 | Check source notice | 10193489352351 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262 | Check source notice | 40193489352352 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262 | Check source notice | 10195327286477 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262 | Check source notice | 40195327286478 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262 | Check source notice | 10198459119477 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262 | Check source notice | 40198459119478 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262 | Check source notice | 10889942624331 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262 | Check source notice | 40889942624332 |
FDA code information
DYNJ0101278F UDI-DI 10193489352351 (EA) 40193489352352 (CS) LOTS 22CDB133 22EDA975 22FDA831 22HDC403 DYNJ0101278G UDI-DI 10195327286477 (EA) 40195327286478 (CS) LOTS 22KDB109 22LDA476 23CDA805 23DDB546 23EDC278 23FDC023 23JDA625 23KDB893 24ADB642 24BDB842 24CDB908 24GDA043 24IDA026 DYNJ0101278I UDI-DI 10198459119477 (EA) 40198459119478 (CS) LOT 24JDC169 DYNJ56262 UDI-DI 10889942624331 (EA) 40889942624332 (CS) LOT 24ABO125 ***Updated 5/20/26 - The following lots were included in the customer letter but not in the FDA submission.*** DYNJ0101278F lots 21ADC024 21CDA355 21EDA527 21GDC529 21IDA040 21IDA786 21JDB214 21KDC443 22ADC117 22GDA789; DYNJ0101278G lots 23HDC100; DYNJ0101278I lots 25AME306 25AMI354 25BMG147 25BMG147A
Why was this product recalled?
Unapproved design changes to the products outside of the 510(k) clearance.
What actions did Medline Industries, LP take?
On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1880-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.