Ongoing
FDA lists this Class II recall for Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1886-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 27, 2026
- FDA report date
- Apr 29, 2026
- Quantity in commerce
- 270,311 total
- Distribution
- See key details
- FDA cause
- Component change control
What is being recalled?
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A
Affected model & UDI checker
Search the exact identifiers published for Z-1886-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A | Check source notice | 10889942227563 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A | Check source notice | 40889942227564 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A | Check source notice | 10195327370756 |
| Not specified | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A | Check source notice | 40195327370757 |
FDA code information
DYNJ35056A UDI-DI 10889942227563 (EA) 40889942227564 (CS) LOT 24ABN639 DYNJ69318A UDI-DI 10195327370756 (EA) 40195327370757 (CS) LOTS 23DMD456 23EMB847 23HMB969 23IMC364 23JME011 23JMI841 23LMC384 24BMD333 24CMB768 24CMJ586 24EMJ527 24GMD970 24JMB408 24JMI594 24KMC774 ***Updated 5/20/26 - The following lots were included in the customer letter but not the FDA submission.*** DYNJ69318A lots 23FMA845 23FMA924 25BMF603
Why was this product recalled?
Unapproved design changes to the products outside of the 510(k) clearance.
What actions did Medline Industries, LP take?
On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1886-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology