Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A Recall

Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1886-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 27, 2026
FDA report date
Apr 29, 2026
Quantity in commerce
270,311 total
Distribution
See key details
FDA cause
Component change control

What is being recalled?

Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318A

Affected model & UDI checker

Search the exact identifiers published for Z-1886-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMedline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318ACheck source notice10889942227563
Not specifiedMedline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318ACheck source notice40889942227564
Not specifiedMedline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318ACheck source notice10195327370756
Not specifiedMedline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EYE PACK DYNJ35056A HEART CATH PACK DYNJ69318ACheck source notice40195327370757
FDA code information

DYNJ35056A UDI-DI 10889942227563 (EA) 40889942227564 (CS) LOT 24ABN639 DYNJ69318A UDI-DI 10195327370756 (EA) 40195327370757 (CS) LOTS 23DMD456 23EMB847 23HMB969 23IMC364 23JME011 23JMI841 23LMC384 24BMD333 24CMB768 24CMJ586 24EMJ527 24GMD970 24JMB408 24JMI594 24KMC774 ***Updated 5/20/26 - The following lots were included in the customer letter but not the FDA submission.*** DYNJ69318A lots 23FMA845 23FMA924 25BMF603

View the original FDA recall notice

Why was this product recalled?

Unapproved design changes to the products outside of the 510(k) clearance.

What actions did Medline Industries, LP take?

On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.