MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3 Recall
MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3
FDA recall Class II
Completed
FDA lists this Class II recall for MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1905-2025
- Recall classification
- Class II
- Enforcement status
- Completed
- Initiation date
- Apr 23, 2025
- FDA report date
- Jun 11, 2025
- Quantity in commerce
- 3 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3
Affected model & UDI checker
Search the exact identifiers published for Z-1905-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MIRA-CAP-110 | MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3 | Check source notice | Not specified |
| Not specified | MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3 | Check source notice | 00850038042028 |
FDA code information
UDI-DI (01)00850038042028; Lot Numbers: 1020040211, 1020040212, 1020040213
Why was this product recalled?
The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did VIRTUAL INCISION CORPORATION take?
A "MEDICAL DEVICE CORRECTION NOTIFICATION" letter dated 4/22/25 was sent to customers. Actions to be taken by the Customer/User: Do not use the device and transition scheduled procedures to another minimally invasive surgical technique i.e. laparoscopic or other robotic assisted device. Remove the device from service using your internal institutional procedures (e.g. Lock-Out, Tag-Out) for device maintenance until Virtual Incision technicians service the devices on hand. There is no anticipated impact to the conduct of the clinical study under the IDE. Please complete the attached Acknowledgement and Receipt form and return to Virtual Incision to indicate you received this letter. Customer Service, Clinical Support Staff, abby.cooper@virtualincision.com,913-608-3214. Quality Assurance, Sr Director of Quality and Manufacturing Engineering, Brandon.williams@virtualincision.com, 308-529-1987. Regulatory Affair, VP of Clinical and Regulatory Affairs, Florence.Beck@virtualincision.com, 352-275-726. Type of Action by the Company: Virtual Incision will be working with your site representatives to schedule on-site servicing of the device with updated software for the Surgeon Control Console within 30 days from receipt of this letter. Immediate corrections have been implemented through the analysis of the system data. Results indicate that implemented software updates have resolved the issue that could contribute to this deficiency in the future. If you have any question, call 513-500-4459 or email: mirasupport@virtualincision.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1905-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology