Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1910-2025

ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample) Recall

ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)

FDA recall Class II

Ongoing

FDA lists this Class II recall for ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1910-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 2, 2025
FDA report date
Jun 11, 2025
Quantity in commerce
193035 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)

Affected model & UDI checker

Search the exact identifiers published for Z-1910-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
UCC-222CPChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)Check source noticeNot specified
FDA code information

Lot Numbers: 200014 200015 200016 900153 900154 900155 900156 900157 900158 900159 900160 900161 900162 900163 900164 900165 900166 900167 900168 900169 900170 900171 900172 900173 900174 900175 900176 900177 103564 103312 103293 102807 102096 101673

View the original FDA recall notice

Why was this product recalled?

During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.

What actions did Healthmark Industries Co., Inc. take?

An "URGENT PRODUCT RECALL" notification letter was sent to customers beginning 5/2/25. Recommended Mitigations/ User Recommendations Please discontinue use of the Pre-Filled Water Syringes containing non-sterile water, which can be identified by a red cap. Any unused product can be returned for credit or replacement. Customer are requested to take the following actions 1. Ensure all affected product in your facility s inventory is no longer in use and has been properly quarantined. 2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected products have been transferred. 3. Contact Customer Service via email at HMCS@hmark.com or 800-521-6224 to return your unused affected product or if you have any questions. 4. Complete and return the response form Getinge by e-mailing a copy to recallresponses.qrc@getinge.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.