Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Recall
Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1914-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 28, 2025
- FDA report date
- Jun 18, 2025
- Quantity in commerce
- Not reported
- Distribution
- See key details
- FDA cause
- Process design
What is being recalled?
Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.
Affected model & UDI checker
Search the exact identifiers published for Z-1914-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 84250320S. | Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. | Check source notice | 07613327479515 |
| 84250320S | Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. | Check source notice | Not specified |
FDA code information
Catalog Number: 84250320S. GTIN: 07613327479515. Lot Number: K1BC9B7.
Why was this product recalled?
Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.
What actions did STRYKER GmbH take?
Stryker notified consignees on about 05/28/2025 via FedEx letter. Consignees were instructed to review inventory and immediately quarantine any affected units on hand, and complete and return the Business Reply Form documenting receipt and product segregation. Upon receipt of the response form, Stryker will contact consignees to arrange for return of the affected units. If affected products have been further distributed, consignees were instructed to notify those customers and let Stryker know.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1914-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology