Rover Mobile X-ray System Recall
Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71
FDA recall Class II
Ongoing
FDA lists this Class II recall for Rover Mobile X-ray System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1918-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 27, 2026
- FDA report date
- Apr 29, 2026
- Quantity in commerce
- 32 units
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71
Affected model & UDI checker
Search the exact identifiers published for Z-1918-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MXU-RV35 | Rover Mobile X-ray System | Check source notice | Not specified |
| MXU-RV71 | Rover Mobile X-ray System | Check source notice | Not specified |
| Not specified | Rover Mobile X-ray System | Check source notice | 09357123000051 |
| Not specified | Rover Mobile X-ray System | Check source notice | 09357123000037 |
FDA code information
Lot Code: Refer to product traceability spreadsheet provided MXU-RV71 UDI: (01)09357123000051/ Serial Numbers: 334 397 399 408 423 425 459 MXU-RV35 UDI: (01)09357123000037/ Serial Numbers: 269 223 349 348 351 363 388 384 391 359 416 419 428 429 430 431 435 434 436 413 437 438 460 461 462
Why was this product recalled?
X-Ray generator may malfunction resulting in the X-Ray being inoperable.
What actions did Micro-X Ltd. take?
On March 27, 2026, Micro-X issued a Urgent: Medical Device Correction Notification, via E-Mail. Micro-X ask consignees to take the following actions: " Contact your local service representative to arrange replacement of the PCBA with the updated version at the next scheduled visit or Preventative Maintenance. In the interim, customers may continue to use the device. " To minimize the likelihood of system failure, the batteries should not be permitted to fully discharge. Any system errors or abnormal behaviours must be reported to your authorised service provider without delay. " If the product has been further distributed to other facilities, please ensure this notice is immediately shared with them. " Please complete and email the Customer Response Form in Appendix 1 to the Micro-X LTD in Australia: djerkovic@micro-x.com and copy support@micro-x.com Completing this form confirms receipt of this notice, understanding of the issue, and the necessary actions to be taken.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1918-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026