Access Total T4 Calibrator, Catalog No. 33805 Recall
Access Total T4 Calibrator, Catalog No. 33805
FDA recall Class II
Ongoing
FDA lists this Class II recall for Access Total T4 Calibrator, Catalog No. 33805. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1922-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 16, 2026
- FDA report date
- Apr 29, 2026
- Quantity in commerce
- 2068 US, 2830 OUS
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Access Total T4 Calibrator, Catalog No. 33805
Affected model & UDI checker
Search the exact identifiers published for Z-1922-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 33805 | Access Total T4 Calibrator, Catalog No. 33805 | Check source notice | Not specified |
| Not specified | Access Total T4 Calibrator, Catalog No. 33805 | Check source notice | 15099590225872 |
FDA code information
UDI-DI 15099590225872 Lots 538729, 538815, 538898, 539107, 539243
Why was this product recalled?
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
What actions did BECKMAN COULTER, INC. take?
Pending
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1922-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology