Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1925-2025

Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Recall

Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital Acquisition Box) Software Version: N/A Product Description: OBM00002 OBM DAB (Digital Acquisition Box) Component: No

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1925-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 7, 2025
FDA report date
Jun 18, 2025
Quantity in commerce
3069
Distribution
See key details
FDA cause
Employee error

What is being recalled?

Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital Acquisition Box) Software Version: N/A Product Description: OBM00002 OBM DAB (Digital Acquisition Box) Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-1925-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
OBM00002Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box)Check source noticeNot specified
Not specifiedBrand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box)Check source notice00382830010825
FDA code information

UDI-DI 00382830010825 Serial Numbers 001 through OBM00002H3613. Note earlier version serial numbers did not contain the sequence of "OBM"

View the original FDA recall notice

Why was this product recalled?

Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.

What actions did Natus Neurology DBA Excel Tech., Ltd. (XLTEK) take?

On May 7, 2025, the firm began notifying customers through "Urgent Medical Device Recall" letters. Customers are instructed to check their Olympic Brainz Monitor stock and determine if they have any devices with the incorrect label orientation. Immediately cease use of incorrectly labeled product and quarantine until you receive return instructions from Natus. Customers with incorrect product labelling will be provided with replacement device(s). If you have any questions regarding this notice, please contact FCA@Natus.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology