Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1939-2025

Femoral Nail, LEFT T2 GTN 8x300 mm. Recall

Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Femoral Nail, LEFT T2 GTN 8x300 mm.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1939-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 22, 2025
FDA report date
Jun 18, 2025
Quantity in commerce
1
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S.

Affected model & UDI checker

Search the exact identifiers published for Z-1939-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1850-0830S.Femoral Nail, LEFT T2 GTN 8x300 mm.Check source notice04546540646293
1850-0830SFemoral Nail, LEFT T2 GTN 8x300 mm.Check source noticeNot specified
FDA code information

Catalog Number: 1850-0830S. GTIN: 04546540646293. Batch lot number: K09F126

View the original FDA recall notice

Why was this product recalled?

Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.

What actions did STRYKER GmbH take?

Stryker notified consignees on about 05/22/2025 via letter. Consignees were instructed to immediately check their inventory for affected units and quarantine any on hand, and complete and return the Business Reply Form by email to . Upon receipt of the completed Business Reply Form, Stryker will contact customers to arrange for the return of their product(s). If consignees have further distributed the product, they were instructed to notify their customers.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology