Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Recall
Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1946-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 30, 2026
- FDA report date
- May 6, 2026
- Quantity in commerce
- 66960
- Distribution
- See key details
- FDA cause
- Vendor change control
What is being recalled?
Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A
Affected model & UDI checker
Search the exact identifiers published for Z-1946-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 00-5049-011-00 | Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula | Check source notice | Not specified |
| 00504901100 | Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula | Check source notice | Not specified |
| Not specified | Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula | Check source notice | 00889024379718 |
| Not specified | Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula | Check source notice | 00889024376564 |
FDA code information
Lot Code: Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301014(10)85390029 ; Individual UDI (01)00889024376564(17)301014(10)85390029 ; Lot Number 85390029 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301103(10)85682882 ; Individual UDI (01)00889024376564(17)301103(10)85682882 ; Lot Number 85682882 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301001(10)85390028 ; Individual UDI (01)00889024376564(17)301001(10)85390028 ; Lot Number 85390028 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300422(10)82395239 ; Individual UDI (01)00889024376564(17)300422(10)82395239 ; Lot Number 82395239 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300423(10)82395240 ; Individual UDI (01)00889024376564(17)300423(10)82395240 ; Lot Number 82395240 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300929(10)85390027 ; Individual UDI (01)00889024376564(17)300929(10)85390027 ; Lot Number 85390027 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300924(10)85390026 ; Individual UDI (01)00889024376564(17)300924(10)85390026 ; Lot Number 85390026 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300415(10)82395238 ; Individual UDI (01)00889024376564(17)300415(10)82395238 ; Lot Number 82395238 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300922(10)85390025 ; Individual UDI (01)00889024376564(17)300922(10)85390025 ; Lot Number 85390025 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300826(10)85390022 ; Individual UDI (01)00889024376564(17)300826(10)85390022 ; Lot Number 85390022 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300918(10)85390024 ; Individual UDI (01)00889024376564(17)300918(10)85390024 ; Lot Number 85390024 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300722(10)85390023 ; Individual UDI (01)00889024376564(17)300722(10)85390023 ; Lot Number 85390023 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)310122(10)86970970 ; Individual UDI (01)00889024376564(17)310122(10)86970970 ; Lot Number 86970970
Why was this product recalled?
Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
What actions did Zimmer, Inc. take?
On March 30, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Risk Managers were instructed to notify affected personnel and locate and quarantine affected product in inventory. A Zimmer Biomet sales representative may remove and return the affected product on your behalf, or the customer may choose to directly return all affected product from their facility. Surgeons were informed that there are no specific patient-monitoring instructions recommended beyond existing follow up schedules.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1946-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology