Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Recall
Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1949-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 23, 2026
- FDA report date
- May 6, 2026
- Quantity in commerce
- 1200 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.
Affected model & UDI checker
Search the exact identifiers published for Z-1949-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| CMP-01669 | Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) | Check source notice | Not specified |
| Not specified | Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) | Check source notice | 15420045514201 |
FDA code information
Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;
Why was this product recalled?
It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.
What actions did Hologic, Inc. take?
On March 23, 2026, Urgent Medical Device Recall letters were sent to customers. Actions to be taken: If you are receiving this notification, our records indicate that you have received a Genius" Review Station Monitor (CMP-01669). As this field correction applies to all monitors, please follow the instructions below and complete the Customer Response Form (CRF) using the link or QR code provided below. If you are a distributor or otherwise supply Genius" Review Station Monitors to downstream customers, you are responsible for promptly forwarding this notification to all affected sub-accounts and ensuring completion of the CRF for each end-user site. Do not change monitor settings " Continue to use the Genius" Review Station only with the original monitor settings configured by Hologic at installation. " Do not: " Disable Barco Application Appearance Manager (AAM) " Change Barco QAWeb calibration or compliance test settings " Adjust or otherwise modify the monitor configuration parameters outside of the original configuration by Hologic at installation. " Only Hologic, or its authorized provider may modify Genius" Review Station monitor configuration settings. If monitor settings have been altered " If you are aware that monitor display or calibration settings have been changed from the original configuration, do not use the Genius" Review Station monitor for clinical review until the monitor has been returned to the original configuration by Hologic or its authorized service provider. " Complete the Customer Response Form (CRF) so Hologic and its field service partner, IQVIA MedTech, can coordinate restoration of the approved configuration. Please promptly fill out the Customer Response Form (CRF) to acknowledge you have received this notification. Distributors should ensure that all downstream customers complete the CRF or provide the required information on their behalf.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1949-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.