Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808 Recall
Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808
FDA recall Class II
Ongoing
FDA lists this Class II recall for Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1963-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 14, 2026
- FDA report date
- May 6, 2026
- Quantity in commerce
- 103 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808
Affected model & UDI checker
Search the exact identifiers published for Z-1963-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808 | Check source notice | 00884838088115 |
| Not specified | Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808 | Check source notice | 00884838098336 |
| Not specified | Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808 | Check source notice | 00884838108608 |
FDA code information
1. Model Number (REF): 781358. UDI: (01)00884838088115(21)45121, (01)00884838088115(21)45148, (01)00884838088115(21)45160, (01)00884838088115(21)45438, (01)00884838088115(21)45034, (01)00884838088115(21)45049, (01)00884838088115(21)45066, (01)00884838088115(21)45090, (01)00884838088115(21)45141, (01)00884838088115(21)45149, (01)00884838088115(21)45170, (01)00884838088115(21)45250, (01)00884838088115(21)45290, (01)00884838088115(21)45085, (01)00884838088115(21)45312, (01)00884838088115(21)45376, (01)00884838088115(21)45122, (01)00884838088115(21)45169, (01)00884838088115(21)45191, (01)00884838088115(21)45192, (01)00884838088115(21)45203, (01)00884838088115(21)45204, (01)00884838088115(21)45207, (01)00884838088115(21)45220, (01)00884838088115(21)45228, (01)00884838088115(21)45229, (01)00884838088115(21)45249, (01)00884838088115(21)45278, (01)00884838088115(21)45279, (01)00884838088115(21)45320, (01)00884838088115(21)45028, (01)00884838088115(21)45195, (01)00884838088115(21)45174, (01)00884838088115(21)45165, (01)00884838088115(21)45382, (01)00884838088115(21)45011, (01)00884838088115(21)45055, (01)00884838088115(21)45075, (01)00884838088115(21)45113, (01)00884838088115(21)45147, (01)00884838088115(21)45152, (01)00884838088115(21)45235, (01)00884838088115(21)45237, (01)00884838088115(21)45245, (01)00884838088115(21)45264, (01)00884838088115(21)45269, (01)00884838088115(21)45277, (01)00884838088115(21)45316, (01)00884838088115(21)45322, (01)00884838088115(21)45337, (01)00884838088115(21)45338, (01)00884838088115(21)45353, (01)00884838088115(21)45398, (01)00884838088115(21)45414, (01)00884838088115(21)45429, (01)00884838088115(21)45051. Serial Numbers: 45121, 45148, 45160, 45438, 45034, 45049, 45066, 45090, 45141, 45149, 45170, 45250, 45290, 45085, 45312, 45376, 45122, 45169, 45191, 45192, 45203, 45204, 45207, 45220, 45228, 45229, 45249, 45278, 45279, 45320, 45028, 45195, 45174, 45165, 45382, 45011, 45055, 45075, 45113, 45147, 45152, 45235, 45237, 45245, 45264, 45269, 45277, 45316, 45322, 45337, 45338, 45353, 45398, 45414, 45429, 45051. 2. Model Number (REF): 782107. UDI: (01)00884838098336(21)45590, (01)00884838098336(21)62158, (01)00884838098336(21)62180, (01)00884838098336(21)62044, (01)00884838098336(21)45567, (01)00884838098336(21)62031, (01)00884838098336(21)62166, (01)00884838098336(21)62184, (01)00884838098336(21)62190, (01)00884838098336(21)62240, (01)00884838098336(21)45519, (01)00884838098336(21)45533, (01)00884838098336(21)62024, (01)00884838098336(21)62047, (01)00884838098336(21)62048, (01)00884838098336(21)62051, (01)00884838098336(21)62052, (01)00884838098336(21)45554, (01)00884838098336(21)62145, (01)00884838098336(21)62146, (01)00884838098336(21)62148, (01)00884838098336(21)45500, (01)00884838098336(21)45501, (01)00884838098336(21)45510, (01)00884838098336(21)45511, (01)00884838098336(21)45514, (01)00884838098336(21)45516, (01)00884838098336(21)45518, (01)00884838098336(21)45589, (01)00884838098336(21)62017, (01)00884838098336(21)62039, (01)00884838098336(21)62073, (01)00884838098336(21)62081, (01)00884838098336(21)62088, (01)00884838098336(21)45521, (01)00884838098336(21)45600, (01)00884838098336(21)45563, (01)00884838098336(21)45570, (01)00884838098336(21)62153, (01)00884838098336(21)62181, (01)00884838098336(21)62182, (01)00884838098336(21)62183. Serial Numbers: 45590, 62158, 62180, 62044, 45567, 62031, 62166, 62184, 62190, 62240, 45519, 45533, 62024, 62047, 62048, 62051, 62052, 45554, 62145, 62146, 62148, 45500, 45501, 45510, 45511, 45514, 45516, 45518, 45589, 62017, 62039, 62073, 62081, 62088, 45521, 45600, 45563, 45570, 62153, 62181, 62182, 62183. 3. Model Number (REF): 782136. UDI: (01)00884838108608(21)28568, (01)00884838108608(21)28503, (01)00884838108608(21)28515, (01)00884838108608(21)28528, (01)00884838108608(21)28577. Serial Numbers: 28568, 28503, 28515, 28528, 28577.
Why was this product recalled?
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Philips North America take?
Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1963-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology