Baxter Novum IQ Syringe Pump, product code 40800BAXUS Recall
Baxter Novum IQ Syringe Pump, product code 40800BAXUS,
FDA recall Class II
Ongoing
FDA lists this Class II recall for Baxter Novum IQ Syringe Pump, product code 40800BAXUS. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1964-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 12, 2025
- FDA report date
- Jun 18, 2025
- Quantity in commerce
- 8027 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Baxter Novum IQ Syringe Pump, product code 40800BAXUS,
Affected model & UDI checker
Search the exact identifiers published for Z-1964-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Baxter Novum IQ Syringe Pump, product code 40800BAXUS | Check source notice | 05413765852428 |
FDA code information
UDI/DI 05413765852428, All Serial Numbers
Why was this product recalled?
Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.
What actions did Baxter Healthcare Corporation take?
Baxter issued an URGENT MEDICAL DEVICE RECALL notice to is consignees on 05/12/2025 via USPS First Class Mail. The notice explained the issue, risk, and requested the following: Actions to be Taken by Customers 1. Customers may continue to use the Novum IQ syringe pump until the pumps can be repaired. If the pump detects a syringe when no syringe is present, or you experience System Error 21502 (Flange Sensor Failure), power cycle the device and reload the syringe. Complete the therapy and remove the pump from service. 2. A Baxter representative will contact your facility to coordinate the repair of your Novum IQ syringe pumps. Please note you will be receiving this repair from Baxter at no charge. 3. Acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 4. Please forward a copy of this communication to the Chief Medical Officer, Medical Director, Director of Nursing, Director of Pharmacy, Facility Risk Manager, Director of Purchasing/Central Supply, and any other departments within your institution who use the affected product. For general questions regarding this communication, or if you experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867, Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1964-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.