Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1965-2025

iBOT PMD with software version 01.05.24. Personal Mobility Device. Recall

iBOT PMD with software version 01.05.24. Personal Mobility Device.

FDA recall Class II

Ongoing

FDA lists this Class II recall for iBOT PMD with software version 01.05.24. Personal Mobility Device.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1965-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 24, 2025
FDA report date
Jun 18, 2025
Quantity in commerce
442 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

iBOT PMD with software version 01.05.24. Personal Mobility Device.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Software Version 01.05.24; UDI (01) 00857584008010 (21) 0001108, (01) 00857584008010 (21) 0001110, (01) 00857584008010 (21) 0001121, (01) 00857584008010 (21) 0001123, (01) 00857584008010 (21) 0001128, (01) 00857584008010 (21) 0001136, (01) 00857584008010 (21) 0001137, (01) 00857584008010 (21) 0001422, (01) 00857584008010 (21) 0001438, (01) 00857584008010 (21) 0001496, (01) 00857584008010 (21) 0001505, (01) 00857584008010 (21) 0001508, (01) 00857584008010 (21) 0001516, (01) 00857584008010 (21) 0001517, (01) 00857584008010 (21) 0001519; Serial No. 0001108, 0001422, 0001438, 0001496, 0001505, 0001508, 0001516, 0001517, 0001519, 0001110, 0001121, 0001123, 0001128, 0001136, 0001137.

View the original FDA recall notice

Why was this product recalled?

Software issue that could potentially lead to the device tipping over from Balance Mode.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Mobius Mobility LLC take?

Consignees were initially contacted via telephone on 4/24/25 to notify them of this software issue. An URGENT MEDICAL DEVICE CORRECTION NOTICE dated 5/5/25 was emailed to consignees to notify them of the recall and instructing them that the firm is developing a software correction. Once Mobius Mobility has obtained a correction they will reach out to consignees to schedule a time to update affected devices. Consignees with any questions can email service @mobiusmobility.com or call 833-346-4268x2.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology