REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile Recall
REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile
FDA recall Class II
Ongoing
FDA lists this Class II recall for REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1977-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 24, 2024
- FDA report date
- Jun 25, 2025
- Quantity in commerce
- 2 units
- Distribution
- See key details
- FDA cause
- Use error
What is being recalled?
REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile
Affected model & UDI checker
Search the exact identifiers published for Z-1977-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 99-709245 | REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile | Check source notice | Not specified |
| D9mm | REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile | Check source notice | Not specified |
| Not specified | REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile | Check source notice | 18059015373653 |
FDA code information
Lot # B4771336 and B4771337, UDI: 18059015373653
Why was this product recalled?
Retrospective; Their is the potential that implantable intramedullary lengthening nails may bend at the distal tail during insertion resulting in having to be replaced.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did ORTHOFIX SRL take?
No official notification was sent to customers. Orthofix representative was present during the first clinical case and both devices were returned as part of the complaint investigation process.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1977-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology