Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02. Recall

Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1977-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Mar 13, 2026
FDA report date
May 13, 2026
Quantity in commerce
6,052 kits
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02.

Affected model & UDI checker

Search the exact identifiers published for Z-1977-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
SLPC34-01Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02.Check source noticeNot specified
SLPC34-02Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02.Check source noticeNot specified
FDA code information

1. Model Number: SLPC34-01. UDI-DI (Kit Lot Numbers): 10809160450963 (1640169, 1641409, 1641410, 1641411, 1641412, 1641413, 1641414, 1641415, 1645684); 10809160459904 (1649576, 1660882, 1660883, 1660884, 1663896, 1665768). 2. Model Number: SLPC34-02. UDI-DI (Kit Lot Numbers): DHR 108916049515 NAV 10809160511213 (1668809, 1668810, 1668812, 1668813, 1668814, 1668815, 1668816).

View the original FDA recall notice

Why was this product recalled?

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did AVID Medical, Inc. take?

Owens and Minor/Avid Medical notified consignees on about 03/13/2026 and then sent an updated letter on about 03/18/2026 with additional kits added to the affected product list. The notice explained the issue was initiated by Medline Industries, risk associated with syringe disconnection and loose connection, and instructed consignees to identify, segregate and quarantine all affected product, and to apply warning labels to affected AVID kits to ensure all users are aware of the actions requested by Medline. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure. All usage guidelines and the Instructions for Use are provided in Medline s recall letter. Consignees were also instructed to complete and return the Recall Response Form as well as to notify customers if affected kits were further distributed to ensure all end-users are appropriately notified and maintain records of effectiveness.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.