Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1980-2025

GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S Recall

GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

FDA recall Class II

Ongoing

FDA lists this Class II recall for GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1980-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 8, 2025
FDA report date
Jun 25, 2025
Quantity in commerce
36 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

Affected model & UDI checker

Search the exact identifiers published for Z-1980-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
RS-00608-X-SGCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-SCheck source noticeNot specified
FDA code information

Serial Numbers: ZL316367, ZL316446, ZL316462, ZL316475, ZL316481, ZL316484, ZL316490, ZL316493, ZL316503, ZL317057, ZL317087, ZL317088, ZL317090, ZL322386, ZL322387, ZL322390, ZL322392, ZL322409, ZL322412, ZL322439, ZL322441, ZL322442, ZL322443, ZL322445, ZL322447, ZL322451, ZL322455, ZL322456, ZL322462, ZL322463, ZL322466, ZL322467, ZL322470, ZL322472, ZL322473, ZL322837.

View the original FDA recall notice

Why was this product recalled?

Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Ohio Medical Corporation take?

Ohio Medical issued a Field Safety Notice to its consignees on 05/08/2025 followed by telephone calls on 05/13/2025. The notice explained the problem with the device, potential risk, and requested the devices be returned for correction. The notice further stated: "The Field Safety Notice is communicated to all impacted customers and is communicated to competent authorities in the affected markets. Receivers should pass this FSN on to all persons with a need to be advised from within your organization or to any organization where the potentially affected products have been supplied. Please consider end users and others within the supply chain in the circulation of this notice."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology