BD GasPak EZ CO2 Pouch System. Recall
BD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic use
FDA recall Class II
Ongoing
FDA lists this Class II recall for BD GasPak EZ CO2 Pouch System.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1982-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 7, 2026
- FDA report date
- May 6, 2026
- Quantity in commerce
- 25,760 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
BD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic use
Affected model & UDI checker
Search the exact identifiers published for Z-1982-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 260684 | BD GasPak EZ CO2 Pouch System. | Check source notice | Not specified |
| Not specified | BD GasPak EZ CO2 Pouch System. | Check source notice | 30382902606841 |
FDA code information
Catalog Nunber: 260684; UDI-DI: 30382902606841; Lot Numbers: 5303157, 5329857, 5329860.
Why was this product recalled?
BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.
What actions did Becton Dickinson & Co. take?
On 04/07/2026, BD sent URGENT: Medical Device Correction letters to customers and distributors. Consignees were instructed that there are no recommendations for retrospective chart review or repeat patient testing. They are to verify incubation atmospheres by incorporating the use of appropriate gas indicators as part of routine laboratory quality control practices. Each BD GasPak EZ Gas Generating Pouch System should be tested periodically for its ability to provide adequate conditions for the growth of target organisms. For CO2 incubation, laboratories may use BD CO2 Indicators (Catalog No. 271055), along with the impacted catalog 260684, to verify sufficient CO2 presence in the environment. If conditions are not confirmed or are suspected to be out of specification, laboratories should repeat incubation using alternative validated methods or established laboratory procedures to ensure organism recovery. Additionally, consignees were instructed to ensure the contents of the notification are read and understood, share this notification with the facility network and forward it to any customers to whom product may have been distributed, and complete and return the Customer Response Form. Distributors were instructed to ensure the notification is read and understood, identify all customers that purchased any affected product, and provide a copy of the attached customer letter to all customers who received the affected lots to advise them of this Product Advisory on BD's behalf. BD is investigating the root cause and will implement appropriate corrective actions to prevent recurrence of this issue. BD GasPak catalog numbers 260684 and 260685 are currently on backorder due to seal integrity challenges identified in some batches following a recent overhaul of supplier manufacturing equipment. BD anticipates resumption of product shipments by the end of March 2026, with full recovery expected by June 2026.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1982-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology