Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1983-2026

BD GasPak EZ Campy Pouch System. Recall

BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use

FDA recall Class II

Ongoing

FDA lists this Class II recall for BD GasPak EZ Campy Pouch System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1983-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 7, 2026
FDA report date
May 6, 2026
Quantity in commerce
2,200 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use

Affected model & UDI checker

Search the exact identifiers published for Z-1983-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
260685BD GasPak EZ Campy Pouch System.Check source noticeNot specified
Not specifiedBD GasPak EZ Campy Pouch System.Check source notice30382902606858
FDA code information

Catalog Number: 260685; UDI-DI: 30382902606858; Lot Number: 5329863.

View the original FDA recall notice

Why was this product recalled?

BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.

What actions did Becton Dickinson & Co. take?

On 04/07/2026, BD sent URGENT: Medical Device Correction letters to customers and distributors. Consignees were instructed that there are no recommendations for retrospective chart review or repeat patient testing. They are to verify incubation atmospheres by incorporating the use of appropriate gas indicators as part of routine laboratory quality control practices. Each BD GasPak EZ Gas Generating Pouch System should be tested periodically for its ability to provide adequate conditions for the growth of target organisms. For CO2 incubation, laboratories may use BD CO2 Indicators (Catalog No. 271055), along with the impacted catalog 260684, to verify sufficient CO2 presence in the environment. If conditions are not confirmed or are suspected to be out of specification, laboratories should repeat incubation using alternative validated methods or established laboratory procedures to ensure organism recovery. Additionally, consignees were instructed to ensure the contents of the notification are read and understood, share this notification with the facility network and forward it to any customers to whom product may have been distributed, and complete and return the Customer Response Form. Distributors were instructed to ensure the notification is read and understood, identify all customers that purchased any affected product, and provide a copy of the attached customer letter to all customers who received the affected lots to advise them of this Product Advisory on BD's behalf. BD is investigating the root cause and will implement appropriate corrective actions to prevent recurrence of this issue. BD GasPak catalog numbers 260684 and 260685 are currently on backorder due to seal integrity challenges identified in some batches following a recent overhaul of supplier manufacturing equipment. BD anticipates resumption of product shipments by the end of March 2026, with full recovery expected by June 2026.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology