Immy, Myco DDR Trident, Bulk Neutralization Buffer B (60 X 30 mL), REF: TBPN67-60 For processing of clinical specimens for Mycobacterium spp. diagnosis Recall
Immy, Myco DDR Trident, Bulk Neutralization Buffer B (60 X 30 mL), REF: TBPN67-60 For processing of clinical specimens for Mycobacterium spp. diagnosis
FDA recall Class II
Ongoing
FDA lists this Class II recall for Immy, Myco DDR Trident, Bulk Neutralization Buffer B (60 X 30 mL), REF: TBPN67-60 For processing of clinical specimens for Mycobacterium spp. diagnosis. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1984-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 26, 2026
- FDA report date
- May 6, 2026
- Quantity in commerce
- 101 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Immy, Myco DDR Trident, Bulk Neutralization Buffer B (60 X 30 mL), REF: TBPN67-60 For processing of clinical specimens for Mycobacterium spp. diagnosis
Affected model & UDI checker
Search the exact identifiers published for Z-1984-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| TBPN67-60 | Immy, Myco DDR Trident, Bulk Neutralization Buffer B (60 X 30 mL), REF: TBPN67-60 For processing of clinical specimens for Mycobacterium spp. diagnosis | Check source notice | Not specified |
| Not specified | Immy, Myco DDR Trident, Bulk Neutralization Buffer B (60 X 30 mL), REF: TBPN67-60 For processing of clinical specimens for Mycobacterium spp. diagnosis | Check source notice | 00816387021091 |
FDA code information
Lot: F5061154/UDI: 00816387021091
Why was this product recalled?
Neutralization Buffer may contain contaminants
What actions did Immuno-Mycologics, Inc take?
On March 25, 2025 Immuno-Mycologics, Inc (IMMY) issued a Urgent Medical Device Recall Notification via E-Mail. IMMY ask consignees to take the following actions: 1. Immediately identify, count, and segregate any kits from this lot you have in your inventory to prevent them from being used or shipped to your customers. 2. As soon as possible, discard your affected inventory. 3. Complete the attached Acknowledgement and Receipt Form even if you do not have any affected stock remaining in your possession. Return the completed form to IMMY using one of the methods below: - Email: customerservice@immy.com, Mail to: Recall #: 1627497-2026-001, Attn: Hunter Conover, IMMY, Inc. 2701 Corporate Centre Dr. Norman, OK USA 73069 3. Ensure relevant staff members are informed of this recall, including relevant clinicians. Clinicians should review all patient test results. 4. If you have supplied any potentially affected product to another organization, please advise that organization of this recall and send them this notification. Please contact us so we can follow up with them. 5. In case product is in transit, display this letter in a prominent place for one month.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1984-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026