Depuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 8 CEM. Part Number: 150450108. Recall
Depuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 8 CEM. Part Number: 150450108.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Depuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 8 CEM. Part Number: 150450108.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1988-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 3, 2026
- FDA report date
- May 6, 2026
- Quantity in commerce
- 3 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Depuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 8 CEM. Part Number: 150450108.
Affected model & UDI checker
Search the exact identifiers published for Z-1988-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Depuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 8 CEM. Part Number: 150450108. | Check source notice | 10603295543022 |
FDA code information
Part Number: 150450108. UDI-DI: 10603295543022. Lot Number: AF1006441. Expiration Date: 12/31/2035.
Why was this product recalled?
The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did DEPUY (IRELAND) take?
Depuy Ireland notified consignees on about 04/03/2026 via email. Consignees were instructed to examine inventory for affected units and immediately quarantine them, arrange for the return of all affected units, and complete and return the provided Business Response Form. Consignees were also instructed to forward the notification letter to anyone within each facility that needs to be informed, notify customers if affected units were further distributed and post a copy of the notification letter in a visible area for awareness.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1988-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology