FDA RECALL RECORD · Z-1992-2025

5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline Recall

5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline, Model Number: 03-5100-7. b. 5008X Standard HDF Bloodline w/CLIC, Model Number: 03-5100-7C. c. 5008X Standard HDF Bloodline w/ Twister, Model Number: 03-5150-2.

FDA recall Class II

Ongoing

FDA lists this Class II recall for 5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1992-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 29, 2025
FDA report date
Jun 25, 2025
Quantity in commerce
Model Number, Cases (Eaches): 03-5100-7, 76 (1,824); 03-5100-7C, 5 (120); 03-5150-2, 8 (192)
Distribution
See key details
FDA cause
Process control

What is being recalled?

5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline, Model Number: 03-5100-7. b. 5008X Standard HDF Bloodline w/CLIC, Model Number: 03-5100-7C. c. 5008X Standard HDF Bloodline w/ Twister, Model Number: 03-5150-2.

Affected model & UDI checker

Search the exact identifiers published for Z-1992-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
03-5100-75008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF BloodlineCheck source noticeNot specified
03-5100-7C5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF BloodlineCheck source noticeNot specified
03-5150-25008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF BloodlineCheck source noticeNot specified
Not specified5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF BloodlineCheck source notice00840861102365
Not specified5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF BloodlineCheck source notice0084086110237
Not specified5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF BloodlineCheck source notice00840861102389
FDA code information

a. Model Number: 03-5100-7; UDI-DI: 00840861102365 (Bag), 10840861102362 (Case); Lot Numbers: 24LR01214, 24SR01010, 25ER01800. b. Model Number: 03-5100-7C; UDI-DI: 0084086110237 (Bag), 10840861102379 (Case); Lot Number: 24LR01215, 24SR01011, 25ER01804. c. Model Number: 03-5150-2; UDI-DI: 00840861102389 (Bag), 10840861102386 (Case); Lot Number: 24LR01216, 24SR01012, 25ER01805.

View the original FDA recall notice

Why was this product recalled?

Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Fresenius Medical Care Holdings, Inc. take?

Fresenius Medical Care notified its single consignee on 04/29/2025 via telephone call. The consignee was made aware of the issue and affected units and instructed to examine their inventory for those affected units. If they had any affected units, they were instructed to remove them from their inventory immediately and provide contact information to arrange for their return. On 05/14/2025, Fresenius Medical Care expanded the recall to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots. On 06/05/2025, Fresenius Medical Care expanded the recall again to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots. A recall notification letter was sent to the consignee on about 06/23/2025. The letter provided the same information as was verbally communicated in notifications.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology