STA - Owren Koller. Recall

STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of the STA-R, STA Compact, STA Satellite and ST art families.

FDA recall Class II

Ongoing

FDA lists this Class II recall for STA - Owren Koller.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1993-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 21, 2025
FDA report date
Jun 25, 2025
Quantity in commerce
939 US ( 2,810 total worldwide)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of the STA-R, STA Compact, STA Satellite and ST art families.

Affected model & UDI checker

Search the exact identifiers published for Z-1993-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
00360STA - Owren Koller.Check source noticeNot specified
Not specifiedSTA - Owren Koller.Check source notice13607450003601
FDA code information

UDI: (01)13607450003601(11)240331(17)260331(10)270902(241)00360. Lot Number: 270902.

View the original FDA recall notice

Why was this product recalled?

Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Diagnostica Stago, Inc. take?

Diagnostica Stago, Inc. notified consignees on 05/21/2025 via email. Consignees were instructed to review inventory for any affected units, stop using the affected lot and destroy it, and complete and return the acknowledgement form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology