LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness. Recall
LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness.
FDA recall Class II
Ongoing
FDA lists this Class II recall for LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1998-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 21, 2025
- FDA report date
- Jul 2, 2025
- Quantity in commerce
- 5 units
- Distribution
- See key details
- FDA cause
- Employee error
What is being recalled?
LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness.
Affected model & UDI checker
Search the exact identifiers published for Z-1998-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 99576-000025 | LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness. | Check source notice | Not specified |
| 21331-000081 | LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness. | Check source notice | Not specified |
| 94576-000005 | LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness. | Check source notice | Not specified |
| Not specified | LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness. | Check source notice | 00883873904565 |
| Not specified | LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness. | Check source notice | 00883873861875 |
| Not specified | LUCAS 2, 3 and 3.1 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness. | Check source notice | 00883873834329 |
FDA code information
Lucus 2 Catalog Number: 99576-000025 UDI-DI code: 00883873904565 Serial Number: 30090111 Lucas 3 Catalog Number: 21331-000081 UDI-DI code: 00883873861875 Serial Numbers: 35172335 35160570 Lucas 3.1 Catalog Number: 94576-000005 UDI-DI code: 00883873834329 Serial Numbers: 3518A164 3518D104 (Scrapped)
Why was this product recalled?
Due to demonstration units of chest compression system being provided to a customer which may potentially be utilized for clinical use.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Jolife AB take?
On 05/21/2025, the firm send via certified mail an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that demonstration LUCAS devices were given to customers for use without the proper labeling. These demonstration units are not maintained in a clinically usable state and are not to be loaned to customers or used in a clinical setting. Customer are instructed to: 1. Immediately check their internal inventory to locate the product listed on the attached business reply form, stop usage, and remove them from their point of use. 2. Return the enclosed business reply form by email to RSRecall@Stryker.com to confirm receipt of this notification. 3. Upon receipt of the completed business reply form, Stryker will contact customers to arrange for the return and replacement of the affected product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within their facility. Keep this communication close to the affected items until replacement. 5. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients appropriately. Any questions or concerns, contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at medtechsup@stryker.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1998-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology