REF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY Recall
REF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY
FDA recall Class II
Ongoing
FDA lists this Class II recall for REF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1999-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 21, 2025
- FDA report date
- Jul 2, 2025
- Quantity in commerce
- 50
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
REF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY
Affected model & UDI checker
Search the exact identifiers published for Z-1999-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 381423 | REF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY | Check source notice | Not specified |
| Not specified | REF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY | Check source notice | 382903814237 |
FDA code information
Lot # 4229661/UDI: 382903814237
Why was this product recalled?
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
What actions did Becton Dickinson Infusion Therapy Systems, Inc. take?
On May 21, 2025, Becton Dickinson Infusion Therapy (BD) issued a "Urgent: Medical Device Recall" Notification to affected consignees FedEx and E-Mail. BD asked consignees to take the following actions: 1. Immediately review your inventory for product listed in the above table. Cease use and destroy any unused affected devices. 2. For devices that have deployed and are currently in use or if an impacted device was previously used, no additional follow-up activities are required for patients that have completed therapy. 3. Share and post this Product Recall letter within your facility network and forward to any customers you may have distributed the product to ensure awareness. Ensure the contents of this Product Recall are read and understood by those within your organization. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1999-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026