epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; Recall

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

FDA recall Class II

Ongoing

FDA lists this Class II recall for epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2005-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 2, 2026
FDA report date
May 6, 2026
Quantity in commerce
71725 (38025 US, 33700 OUS)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Affected model & UDI checker

Search the exact identifiers published for Z-2005-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;Check source notice00809708121860
FDA code information

Siemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);

View the original FDA recall notice

Why was this product recalled?

Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Healthcare Diagnostics, Inc. take?

On April 2, 2026, Urgent Medical Device Recall letters were sent to customers. Actions to be taken: Perform the instructions provided below: " Discontinue use of and discard the kit lots listed in the table above (Products Section). " Please review this letter with your Medical Director/POC Coordinator to determine the appropriate course of action, including for any previously generated results, if applicable. " Complete and return the Field Correction Effectiveness Check Form and Product Replacement Form attached to this letter within 30 days. " Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. " For any questions, you may call Siemens Healthineers Technical Solutions Center at 877- 229-3711, available daily from 7:00 AM to 11:00 PM EST.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology