epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; Recall
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
FDA recall Class II
Ongoing
FDA lists this Class II recall for epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2005-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 2, 2026
- FDA report date
- May 6, 2026
- Quantity in commerce
- 71725 (38025 US, 33700 OUS)
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
Affected model & UDI checker
Search the exact identifiers published for Z-2005-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; | Check source notice | 00809708121860 |
FDA code information
Siemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);
Why was this product recalled?
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Siemens Healthcare Diagnostics, Inc. take?
On April 2, 2026, Urgent Medical Device Recall letters were sent to customers. Actions to be taken: Perform the instructions provided below: " Discontinue use of and discard the kit lots listed in the table above (Products Section). " Please review this letter with your Medical Director/POC Coordinator to determine the appropriate course of action, including for any previously generated results, if applicable. " Complete and return the Field Correction Effectiveness Check Form and Product Replacement Form attached to this letter within 30 days. " Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. " For any questions, you may call Siemens Healthineers Technical Solutions Center at 877- 229-3711, available daily from 7:00 AM to 11:00 PM EST.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2005-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology