Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2006-2026

Brand Name: Leksell Vantage Arc System REF: 1053958 Recall

Brand Name: Leksell Vantage Arc System REF: 1053958

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Leksell Vantage Arc System REF: 1053958. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2006-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 19, 2026
FDA report date
May 6, 2026
Quantity in commerce
768 (364Left/364Right)
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Brand Name: Leksell Vantage Arc System REF: 1053958

Affected model & UDI checker

Search the exact identifiers published for Z-2006-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1053958Brand Name: Leksell Vantage Arc System REF: 1053958Check source noticeNot specified
Not specifiedBrand Name: Leksell Vantage Arc System REF: 1053958Check source notice07340048304856
Not specifiedBrand Name: Leksell Vantage Arc System REF: 1053958Check source notice07340048304818
Not specifiedBrand Name: Leksell Vantage Arc System REF: 1053958Check source notice07340048304801
FDA code information

Lot Code: Arc System: UDI:07340048304856 Parts of Arc System: Left Z-Slide: Serial Numbers :UDI: 07340048304818 00001 00625 Right Z-Slide: UDI: 07340048304801/ Serial Numbers: 00001 00675

View the original FDA recall notice

Why was this product recalled?

Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.

What actions did Elekta, Inc. take?

On March 19, 2026 Elekta issued a "Urgent: Medical Device Recall" notification via e-mail. Elekta asked consignees 1. Before each operation, products should be cleaned, sterilized (unless specified otherwise), and checked for proper functionality. 2. Make sure to examine all scales and angles to make sure that they are correctly set and securely fixed before proceeding. 3. Make sure to obtain a second person in the OR room to read the coordinate settings so as to ensure they are correct and check they are securely locked. 4. Before you continue, check that all parts are assembled correctly, firmly attached, and locked in place. 5. Inspect the equipment for visible signs of damage before clinical use. Do not use damaged equipment for any purpose. If you suspect that any part is damaged, contact your Elekta representative immediately. 6. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. 7. Advise the appropriate personnel working with this product, on the content of this letter. If the product has been transferred to another facility, please provide them with a copy of this communication.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology