FDA RECALL RECORD · Z-2013-2025

Servo-u Ventilator System. Recall

Servo-u Ventilator System. Model Number: 6688600.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Servo-u Ventilator System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2013-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 30, 2025
FDA report date
Jul 9, 2025
Quantity in commerce
874 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Servo-u Ventilator System. Model Number: 6688600.

Affected model & UDI checker

Search the exact identifiers published for Z-2013-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
6688600Servo-u Ventilator System.Check source noticeNot specified
Not specifiedServo-u Ventilator System.Check source notice07325710001042
FDA code information

Model Number: 6688600. UDI-DI: 07325710001042. All software versions.

View the original FDA recall notice

Why was this product recalled?

Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Maquet Critical Care AB take?

Maquet Critical Care AB, a subsidiary of Getinge, notified consignees on about 05/30/2025 via FedEx letter. Consignees were instructed to configure all affected Servo ventilators with "Extended leakage test" set to "Off", per chapter 8.2.4 Configuration in the User's manuals and all users are recommended to follow the updated instructions for use of the User's manuals chapter 9.5.1 Patient circuit test, which is provided. Consignees are instructed to ensure all devices in inventory have been properly addressed as described above, to ensure the message is forwarded to all affected personnel and to individuals if units were transferred. Additionally, consignees were requested to complete and return the provided response form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.