Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2022-2025

interventional fluoroscopic x-ray system Recall

interventional fluoroscopic x-ray system

FDA recall Class II

Ongoing

FDA lists this Class II recall for interventional fluoroscopic x-ray system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2022-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 15, 2025
FDA report date
Jul 9, 2025
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

interventional fluoroscopic x-ray system

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

4056869149325

View the original FDA recall notice

Why was this product recalled?

In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.

What actions did Siemens Medical Solutions USA, Inc. take?

Siemens Healthineers sent customers a Customer Advisory Notice dated May 15, 2025 informing customers of a potential issue with ARTIS pheno and ARTIS icono systems with the following model numbers: ARTIS pheno 10849000, ARTIS icono biplane 11327600, ARTIS icono ceiling 11328100, ARTIS icono floor 11327700; due to in very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose and appropriate software update will be implemented, based on the current software version of the affected system.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology