FDA RECALL RECORD · Z-2023-2025

BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA). Recall

BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA).

FDA recall Class II

Ongoing

FDA lists this Class II recall for BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA).. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2023-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 20, 2025
FDA report date
Jul 2, 2025
Quantity in commerce
12,256 eaches
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA).

Affected model & UDI checker

Search the exact identifiers published for Z-2023-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
245128BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA).Check source noticeNot specified
FDA code information

Catalog No. 245128; UDI-DI: (01) 0038290245128; Lot No. 4150998, 4177908, 4178500, 4262031, 4262036, 4262037, 4262039, 4262040, 4262044, 4284438, 4284441, 4284443, 4284445, 4284449, 4304789, 4304796, 4347228, 4347230, 4352110, 4362418, 5007456, 5020153, 5050670, 5050671, 5050672, 5050675, 5050678.

View the original FDA recall notice

Why was this product recalled?

BD has received additional complaints of intermittent false resistance results for PZA during susceptibility testing of Mycobacterium tuberculosis isolates.

What actions did Becton Dickinson & Co. take?

***UPDATE 11/13/2025*** BD Issued an URGENT: Medical Device Correction letters were sent to customers. BD is pleased to announce that we have resumed production of a modified version of the BD BACTEC MGIT 960 PZA Kit which includes modified inoculation methods and reduced shelf-life. BD has conducted a thorough review of the performance from all supported inoculum sources using the synthetic raw material and the following change has been implemented for the BD BACTEC MGIT 960 PZA Kit. 1. Only inoculum prepared from MGIT tubes 3-5 days post-positivity can be currently supported. 2. Inoculum prepared from MGIT tubes 1-2 days past instrument positivity are currently not supported. 3. Inoculum prepared from solid media are currently not supported. Additionally, the shelf life of the product has been adjusted from 18 months to 13 months to reflect the latest internal supporting data. Clinical users should refer to the updated Instructions for Use (IFU) available at https://www.bd.qarad.eifu.online/hcp. Additional languages will be added as translations are completed. An URGENT: Medical Device Correction notice dated 5/20/25 was sent to consignees via mail. The notification instructs consignees to inspect their inventory for recalled units and destroy them per their institution's process of destruction. The recall notification is to be shared at consignee locations with those who utilize the device and to customers who received distributed product. Consignees are to return a completed Customer Response Form to BD. Consignees with any questions are to email BDRC12@bd.com. Consignees are to report any complaints by phone to 1-844-823-5433 or by email at productcomplaints@bd.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.