Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2026-2025

Jewel Precision Reusable Rigid Sterilization Container System. Recall

Jewel Precision Reusable Rigid Sterilization Container System. Model/Catalog Number: JP-24-6.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Jewel Precision Reusable Rigid Sterilization Container System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2026-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 22, 2025
FDA report date
Jul 2, 2025
Quantity in commerce
11 units
Distribution
See key details
FDA cause
No Marketing Application

What is being recalled?

Jewel Precision Reusable Rigid Sterilization Container System. Model/Catalog Number: JP-24-6.

Affected model & UDI checker

Search the exact identifiers published for Z-2026-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
JP-24-6Jewel Precision Reusable Rigid Sterilization Container System.Check source noticeNot specified
Not specifiedJewel Precision Reusable Rigid Sterilization Container System.Check source notice00850043393016
FDA code information

Model No JP-24-6. GTIN: 00850043393016. All lots shipped up to May 23, 2025.

View the original FDA recall notice

Why was this product recalled?

The Instructions for Use do not match the 510(k) FDA cleared indications for use or sterility shelf life.

What actions did Jewel Precision Sheet Metal & Machining Co, Inc. take?

Jewel Precision notified consignees on about 05/22/2025 via email. Consignees were instructed to acknowledge the notification, inform Jewel Precision of their distribution of the products, obtain old Instructions for Use from their customers, return any obsolete Instructions for Use to Jewel Precision, and send updated Instructions for Use to their customers.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.