Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2029-2025

GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system Recall

GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2029-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 16, 2025
FDA report date
Jul 2, 2025
Quantity in commerce
34 units
Distribution
See key details
FDA cause
Software design (manufacturing process)

What is being recalled?

GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system

Affected model & UDI checker

Search the exact identifiers published for Z-2029-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5943263GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound systemCheck source noticeNot specified
Not specifiedGE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound systemCheck source notice00195278722430
FDA code information

UDI/DI 00195278722430, Lot/Serial Numbers: 6310028WX0, 6310008WX0, 6310041WX0, 6310017WX0, 6310031WX0, 6310044WX0, 6310024WX0, 6310038WX0, 6310040WX0, 6310019WX0, 6310018WX0, 6310023WX0, 6310033WX0, 6310012WX0, 6310026WX0, 6310020WX0, 6310032WX0, 6310014WX0, 6310021WX0, 6310025WX0, 6310010WX0, 6310027WX0, 6310037WX0, 6310009WX0, 6310007WX0, 6310035WX0, 6310016WX0, 6310039WX0, 6310036WX0, 6310043WX0, 6310011WX0, 6310015WX0, 6310013WX0, 6310029WX0.

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Medical Systems China Co., Ltd. take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 05/16/2025 via letter delivered using a traceable means. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by Customer/User: You may continue to use your Versana Premier R3 and LOGIQ F R3 series ultrasound systems by following the actions below: 1) If you use the Whizz report feature, reboot the system after finishing each obstetric patient examination. OR 2) Instead of using the Whizz report feature, after completing an obstetric patient examination, print all obstetric measurement and calculation results from the worksheet directly. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.71082@gehealthcare.com." If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology