Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226. Recall
Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2032-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- May 30, 2025
- FDA report date
- Jul 16, 2025
- Quantity in commerce
- 9839 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
All LikoScale Adapter Kits manufactured between 8/27/2013 to 2/27/2025
Why was this product recalled?
The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Baxter Healthcare Corporation take?
Baxter issued an URGENT PRODUCT RECALL notice to is consignees on 05/30/2025 via USPS first class mail. The notice described the problem with the device, hazard, and requested the following: "Actions to be taken by customers 1. Please locate and stop using all Q-link 13 components on any of the products listed in the affected product table. 2. Contact Baxter Technical Support to discuss alternative options, if necessary, at 800-445-3720, option 2, between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or by email at HRC_Technical_Support@Baxter.com. 3. Please post this letter in areas where affected mobile lifts are stored and used. 4. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 6. Please forward a copy of this communication to all departments within your institution that use the affected mobile lifts. 7. If you are a dealer, wholesaler or distributor/reseller that distributed any affected product to other facilities, please forward this communication to your customers and check the associated box on the customer portal. Further Information and Support For general questions regarding this communication, please contact Baxter Technical Support at 800-445-3720 Option 2 between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or by email at HRC_Technical_Sup
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2032-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology