LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5 Recall
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
FDA recall Class II
Ongoing
FDA lists this Class II recall for LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2048-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 30, 2026
- FDA report date
- May 13, 2026
- Quantity in commerce
- 2 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
Affected model & UDI checker
Search the exact identifiers published for Z-2048-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| M302 | LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5 | Check source notice | Not specified |
| 6460 | LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5 | Check source notice | Not specified |
| Not specified | LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5 | Check source notice | 00802526620713 |
FDA code information
LUX-DX II UDI-DI: 00802526620713; SERVER SW LATITUDE DRAGON US UDI-DI: SERVER SW LATITUDE DRAGON US; Serial Number: 104263 109787
Why was this product recalled?
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
What actions did BOSTON SCIENTIFIC CORPORATION take?
An IMPORTANT Medical Device Correction notification letter dated 3/30/26 was sent to customers. Recommendations As noted previously, Boston Scientific released an update to LATITUDE Clarity on March 27, 2026 to correct this programming discrepancy. Devices impacted by this issue will be automatically corrected to the clinician assigned programming once the device successfully reconnects with LATITUDE Clarity. To confirm updated programming settings have been applied to the impacted M302/M312 device: " Navigate to the patient s profile on LATITUDE Clarity; " Verify that the last connection to the device occurred after March 27, 2026. If not, the patient will need to reconnect their mobile monitor; " Check device programming under Counters and Settings and verify that Brady, Pause, and PVC Burden are set as desired; " Complete and return the enclosed Acknowledgment Form per the instructions. Boston Scientific s RhythmCARE" Support Team is available to assist HCPs with resolving this programming discrepancy, confirming programmed settings, or assisting with reconnecting a patient, either remotely or in-clinic. Additional Information Patient safety remains Boston Scientific s highest priority. We recognize the impact of this device behavior on both you and your patients, and we are committed to transparent communication with physicians and other HCPs to ensure you have relevant information for managing your patients. If you have additional questions regarding the information described above or would like to report a clinical event, please contact the Boston Scientific RhythmCARE Support team. Adverse events may also be reported directly to the FDA s MedWatch Adverse Event Reporting program. United States RhythmCARE Support 1.800.CARDIAC (227.3422) latitude@bsci.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2048-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology