FDA RECALL RECORD · Z-2064-2025

Philips X5-1 Transducer. Recall

Philips X5-1 Transducer.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips X5-1 Transducer.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2064-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 28, 2025
FDA report date
Jul 9, 2025
Quantity in commerce
5,230 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Philips X5-1 Transducer.

Affected model & UDI checker

Search the exact identifiers published for Z-2064-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
989605400802Philips X5-1 Transducer.Check source noticeNot specified
Not specifiedPhilips X5-1 Transducer.Check source notice00884838061422
FDA code information

Model No. 989605400802; UDI: (01)00884838061422(21)B2XMDQ, (01)00884838061422(21)B3H27D, (01)00884838061422(21)B2ZXZB, (01)00884838061422(21)B26JC4, (01)00884838061422(21)B3GYXN, (01)00884838061422(21)B2ZXD8, (01)00884838061422(21)B3GJB2, (01)00884838061422(21)B2MQZ3, (01)00884838061422(21)B3L198, (01)00884838061422(21)B3KC2D, (01)00884838061422(21)B3MTN1, (01)00884838061422(21)B2894C; Serial No. B2XMDQ, B3H27D, B2ZXZB, B26JC4, B3GYXN, B2ZXD8, B1CGYC, B3GJB2, B2MQZ3, B3L198, B3KC2D, B3MTN1, B2894C, B1FHTZ.

View the original FDA recall notice

Why was this product recalled?

Ultrasound transducer devices were refurbished beyond their useful life.

What actions did Philips Ultrasound, Inc take?

An URGENT Medical Device Recall notification dated 5/28/25 was mailed to consignees. The notification informs consignees that Philips Ultrasound will be replacing transducers that were refurbished beyond their useful life with a new transducer. Consignees may continue to use affected devices while awaiting replacement; consignees should use devices in accordance with their intended use and per the instructions provided int he User Manual. Users should perform visual inspection and electrical safety testing as indicated in the User Manual; if a device does not pass inspection, it should be removed from clinical use. Consignees are asked to return the completed Response Form and circulate the provided notification to all users to ensure awareness. A copy of the recall notification should be kept with affected transducers until their replacement. Consignees with any questions can contact the Customer Care Solutions Center at 1-800-722-9377. Additional consignees were identified and notified via URGENT Medical Device Recall Notification on 9/5/25.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.