Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2066-2026

Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F Recall

Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F

FDA recall Class I

Ongoing

FDA lists this Class I recall for Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2066-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Apr 2, 2026
FDA report date
May 20, 2026
Quantity in commerce
150 kits (US only)
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Pack AMS6908E UDI-DI B098AMS6908E0 Lot 214875 Pack AMS6908F UDI-DI B098AMS6908F0 Lots 215436, 225001, 226706, 234088

View the original FDA recall notice

Why was this product recalled?

Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Windstone Medical Packaging, Inc. take?

Firm began notifying consignees via email on April 2, 2026. Notice was titled "URGENT: MEDICAL DEVICE RECALL." Consignees were instructed to check inventory and quarantine affected products. Firm instructed consignees to over-label affected products with a sticker which states that at the time the kit is opened, the affected Medline syringe should be identified and set aside, and that the affected syringe should be rendered unusable and disposed of. Firm will provide labels for any kits that need to be identified.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.