Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2068-2026

Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.; Recall

Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2068-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 7, 2026
FDA report date
May 13, 2026
Quantity in commerce
101 systems
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;

Affected model & UDI checker

Search the exact identifiers published for Z-2068-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMint Lesion; Software Versions: 3.4.0 up to 3.9.5.;Check source notice04260495880341
Not specifiedMint Lesion; Software Versions: 3.4.0 up to 3.9.5.;Check source notice04260495880358
Not specifiedMint Lesion; Software Versions: 3.4.0 up to 3.9.5.;Check source notice04260495880365
Not specifiedMint Lesion; Software Versions: 3.4.0 up to 3.9.5.;Check source notice04260495880372
Not specifiedMint Lesion; Software Versions: 3.4.0 up to 3.9.5.;Check source notice04260495880389
Not specifiedMint Lesion; Software Versions: 3.4.0 up to 3.9.5.;Check source notice04260495880396
FDA code information

Software Versions: 3.4.0 up to 3.9.5. UDI-DI: 04260495880341, 04260495880358, 04260495880365, 04260495880372, 04260495880389, 04260495880396.

View the original FDA recall notice

Why was this product recalled?

If the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly linked to other patients.

What actions did Mint Medical GmbH take?

On or about January 7, 2026 Urgent Field Safety Notice letters were emailed to customers. Customers were instructed to read the notice and check if any affected product is available for use. If that is the case, the malfunction may have occurred and may occur in your system. Please be aware that the malfunction may occur. Your IT department must ensure that no temporary network interruptions occur in your clinical network. Also, your IT department must ensure that the mint Lesion server is not restarted while users work with mint Lesion application (i.e., the server may be restartet, but only outside of working hours). Please verify correctness of the information of prior time-points. Be attentive to potential inconsistencies in lesion attributes when opening existing cases for follow up. Cross check lesion location and categorization with prior radiological reports and image data. If you suspect that the problem may have occurred on your mint Lesion system, please contact Mint Medical support via support@mint-medical.com or any other contact details for technical analysis and investigation. Mint Medical Support will analyze if the malfunction appeared at your mint Lesion system in the past. If you believe that this failure has occurred, please review the potentially affected radiological reports in your reporting application and take the necessary steps to correct them. Pass this notice on to all users of mint Lesion" within your organization. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology