Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2073-2025

VITROS Chemistry Products OP Reagent Gen 15. Recall

VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect opiates in urine. One sales unit contains 6 Microtip packs of reagent. Each pack has 50 OP tests.

FDA recall Class II

Ongoing

FDA lists this Class II recall for VITROS Chemistry Products OP Reagent Gen 15.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2073-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 12, 2025
FDA report date
Jul 9, 2025
Quantity in commerce
2,017 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect opiates in urine. One sales unit contains 6 Microtip packs of reagent. Each pack has 50 OP tests.

Affected model & UDI checker

Search the exact identifiers published for Z-2073-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
680VITROS Chemistry Products OP Reagent Gen 15.Check source noticeNot specified
Not specifiedVITROS Chemistry Products OP Reagent Gen 15.Check source notice10758750001835
FDA code information

Model Number: 680 1997. UDI-DI: 10758750001835. Lot Numbers: 1527-15-2749 (exp. 17-Nov-2025), 1527-15-2771 (exp. 17-Dec-2025), 1527-15-2818 (exp. 11-Apr-2026), 1527-15-2871 (exp. 15-May-2026).

View the original FDA recall notice

Why was this product recalled?

Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did ORTHO-CLINICAL DIAGNOSTICS, INC. take?

Quidel Ortho notified consignees on about 06/12/2025 via FedEx letter. Distributors were instructed to discontinue distribution and discard inventory of any affected lots, send the provided customer letter and Confirmation of Receipt to any affected customers, and complete and return their Confirmation of Receipt. Customers were instructed that if their laboratory is unable to generate QC results within acceptable range using lots from GEN 15 of VITROS OP Reagent, discontinue using, render unusable, and discard your inventory of VITROS OP Reagent, GEN 15. Additionally, there were requested to complete and return the Confirmation of Receipt, maintain the notification with all records, and provide the notification if affected units were further distributed.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.