Baxter SIGMA Spectrum Infusion Pump V6, Model Number 35700BAX Recall
Baxter SIGMA Spectrum Infusion Pump V6, Model Number 35700BAX
FDA recall Class I
Ongoing
FDA lists this Class I recall for Baxter SIGMA Spectrum Infusion Pump V6, Model Number 35700BAX. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2095-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Jun 20, 2025
- FDA report date
- Jul 23, 2025
- Quantity in commerce
- 14 units
- Distribution
- See key details
- FDA cause
- Employee error
What is being recalled?
Baxter SIGMA Spectrum Infusion Pump V6, Model Number 35700BAX
Affected model & UDI checker
Search the exact identifiers published for Z-2095-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 35700BAX | Baxter SIGMA Spectrum Infusion Pump V6, Model Number 35700BAX | Check source notice | Not specified |
| Not specified | Baxter SIGMA Spectrum Infusion Pump V6, Model Number 35700BAX | Check source notice | 00085412091570 |
FDA code information
UDI-DI: 00085412091570; Serial Numbers: 765100 , 772744 , 789742 , 841197 , 867753 , 920606 , 1024015 770164 788724 , 793079 , 842498 , 918342 , 962810 , 1027523
Why was this product recalled?
Certain Spectrum infusion pumps may have an incorrect version of software.
What actions did Baxter Healthcare Corporation take?
An URGENT MEDICAL DEVICE RECALL notification letter dated 6/20/25 was sent to customers. Actions to be Taken by Customers 1. Immediately locate Spectrum pumps with the affected serial numbers and remove them from service. The product code and serial number can be found on the bottom of the infusion pump. 2. Check and record the software version installed on the pump. The software on a V6 pump should begin with the number 6 and the software on a V8 pump should begin with the number 8. The software version is located on the startup screen upon powering up the pump as pictured on the next page. Additionally, the software version is indicated in the pump menu which can be accessed by following the steps. For V6: From the Pump Information screen, press the sw info soft key to display the software version screen. For V8: From the Select Care Area screen, press the options menu soft key. Select User Options and press OK. Select View Information and press OK. Select Pump Information and press OK to display the Pump Info screen. 3. Contact Baxter to confirm the software version on the affected pumps and to schedule service if required. Baxter Global Technical Services can be reached at 800-843-7867 (select option 1, then option 2), Monday through Friday, between 7:00 am and 7:00 pm Eastern Time. 4. If you received this communication directly from Baxter, acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2095-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology