FDA RECALL RECORD · Z-2101-2025

O-arm O2 Imaging System. Mobile X-Ray System. Recall

O-arm O2 Imaging System. Mobile X-Ray System.

FDA recall Class II

Ongoing

FDA lists this Class II recall for O-arm O2 Imaging System. Mobile X-Ray System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2101-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 27, 2025
FDA report date
Jul 16, 2025
Quantity in commerce
3 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

O-arm O2 Imaging System. Mobile X-Ray System.

Affected model & UDI checker

Search the exact identifiers published for Z-2101-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedO-arm O2 Imaging System. Mobile X-Ray System.Check source notice00763000709631
FDA code information

CFN: BI70002000; UDI: 00763000709631; Serial No. C4014, C4017, C4019.

View the original FDA recall notice

Why was this product recalled?

A mechanical component that supports the O2 gantry to the support structure in affected devices may have been damaged during manufacture.

What actions did Medtronic Navigation, Inc.-Boxborough take?

An URGENT: MEDICAL DEVICE CORRECTION notice dated 5/27/25 was mailed to consignees. In the notice consignees are instructed to discontinue use of affected systems until it can be serviced by Medtronic personnel; a copy of the recall notification is to be posted with the system until the issue has been resolved. A Medtronic Field Service Engineer will contact consignees to schedule a visit to inspect affected systems and implement any corrections needed. Consignees are to complete and return the provided Customer Confirmation Form and forward the recall notification to all users and facilities to which product was further distributed. Consignees with any questions can contact Technical Services for help by phone at 1-888-826-5603 or by email at rs.navtechsupport@medtronic.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.