Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2102-2025

i-view video laryngoscope. Recall

i-view video laryngoscope. Model Number: 8008000. The i-view video laryngoscope is indicated to facilitate laryngoscopy.

FDA recall Class I

Ongoing

FDA lists this Class I recall for i-view video laryngoscope.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2102-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jun 17, 2025
FDA report date
Jul 23, 2025
Quantity in commerce
2,152 pieces
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

i-view video laryngoscope. Model Number: 8008000. The i-view video laryngoscope is indicated to facilitate laryngoscopy.

Affected model & UDI checker

Search the exact identifiers published for Z-2102-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
8008000i-view video laryngoscope.Check source noticeNot specified
FDA code information

Model Number: 8008000. UDI-DI (Case): 05030267166497. UDI-DI (Single Unit): 5030267150649. Lot Numbers: 1240555, 1240793.

View the original FDA recall notice

Why was this product recalled?

Potential for faulty devices as a result of depleted batteries.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Intersurgical Inc take?

Intersurgical notified consignees and account executives via email on 06/17/2025. The notification instructed consignees and account executives to immediately discontinue use and quarantine any affected stock on hand, notify customers if affected units have been further distributed, and contact Intersurgical Customer Support at 800-828-9633 for a Return Goods Authorization (RGA) Number and return label to return all affected units. Also, they were instructed to complete and return the Customer Product Recall Acknowledgement Form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology