Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2134-2025

Azurion 7 M20. Fluoroscopic X-Ray System. Recall

Azurion 7 M20. Fluoroscopic X-Ray System.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Azurion 7 M20. Fluoroscopic X-Ray System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2134-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 4, 2025
FDA report date
Jul 30, 2025
Quantity in commerce
78 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Azurion 7 M20. Fluoroscopic X-Ray System.

Affected model & UDI checker

Search the exact identifiers published for Z-2134-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
722079Azurion 7 M20. Fluoroscopic X-Ray System.Check source noticeNot specified
Not specifiedAzurion 7 M20. Fluoroscopic X-Ray System.Check source notice00884838085268
FDA code information

System Model No. 722079; UDI: 00884838085268; Serial No. 525, 584, 587, 613, 631, 658, 708, 774, 777, 793, 800, 801, 838, 842, 843, 862, 865, 866, 879, 881, 896, 899, 900, 905, 908, 909, 912, 915, 923, 925, 930, 939, 948, 955, 957, 992, 750, 814, 321, 370, 549, 567, 670, 687, 688, 693, 707, 743, 759, 767, 768, 775, 785, 786, 791, 819, 855, 871, 872, 902, 916, 917, 933, 936, 956, 960, 963, 965, 966, 969, 977, 993.

View the original FDA recall notice

Why was this product recalled?

Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?

A "Device URGENT Medical Device Correction" notification dated 6/03/2025 was mailed to consignees. Consignees are instructed to circulate the provided notification to all system users and keep a copy of the recall notification with affected devices to maintain awareness. If you notice unusual noise during C-arm roll motion, or if the C-arm roll motion is disabled and the user message Motorized movement is not available appears, immediately contact a Philips representative. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. Affected systems may continue to be used in accordance with their Instructions for Use (IFU). Complete and return the response form included in this Urgent Medical Device letter. Consignees are provided instructions to avoid rapidly pressing and releasing footswitch pedals and how to restart systems that experience the failure mode. The firm will deploy a software update in all affected system by June 2025. Consignees with any questions should contact their local Philips representative at 1-800-722-9377.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology