Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2136-2026

Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Recall

Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: No Product Description: 11-1424 - ICEMAN W/,UNIV,LOOP,NS,EH The IceMan CLASSIC3¿ cold therapy unit helps reduce pain and swelling, speeding up rehabilitation. The IceMan¿ helps provide extended cold therapy to aid in a variety of indications and protocols as directed by a medical professional. The IceMan CLASSIC3 utilizes DonJoy's recirculation technology. Component: No

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2136-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 19, 2026
FDA report date
May 20, 2026
Quantity in commerce
480
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: No Product Description: 11-1424 - ICEMAN W/,UNIV,LOOP,NS,EH The IceMan CLASSIC3 cold therapy unit helps reduce pain and swelling, speeding up rehabilitation. The IceMan helps provide extended cold therapy to aid in a variety of indications and protocols as directed by a medical professional. The IceMan CLASSIC3 utilizes DonJoy's recirculation technology. Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-2136-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
11-1424Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EHCheck source noticeNot specified
Not specifiedBrand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EHCheck source notice00888912017305
FDA code information

Model/Catalog Number: 11-1424 Lot Code: Lot: 110525 GTIN: 00888912017305

View the original FDA recall notice

Why was this product recalled?

Due to broken connector which may result in a temporary interruption of prescribed cold therapy.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Djo, LLC take?

On 03/19/2026, the firm emailed an "Urgent: Medical Device Recall" Letter to customers to inform them that, twenty-nine complaints have been received reporting a consistent failure mode described as a broken connector. The connector is the component that attaches the end of the circulating hose from the cooler to the pad which is placed on/around the affected limb. Excess water leakage from this component is a key indicator of breakage. Customers are instructed to: Enovis (DJO, LLC) requires customers to take the following actions by 03-31-2026: This notice must be forwarded to all those within their organization or to any organization where the affected product has been transferred. 1. Review stock of the affected products and lot numbers to be returned. Quantities or affected product previously shipped are listed in the table at the opening of this letter. 2. Quarantine these devices from sale. 3. Complete the attached acknowledgment and response form and return to (Customer.Care@enovis.com). 4. To return affected product, please contact (Customer.Care@enovis.com) or call (800) 336-6569 option 4 and customers will be provided with a shipping label to return the affected product. 5. Return all affected products using the shipping label provided. For questions - contact Kenneth Newman at productsafety@enovis.com or call/text 682-299-9133.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology