Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2138-2025

Natus Brain Monitor Breakout Box, Part Number 021911; full-montage standard electroencephalograph Recall

Natus Brain Monitor Breakout Box, Part Number 021911; full-montage standard electroencephalograph

FDA recall Class II

Ongoing

FDA lists this Class II recall for Natus Brain Monitor Breakout Box, Part Number 021911; full-montage standard electroencephalograph. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2138-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 18, 2025
FDA report date
Jul 30, 2025
Quantity in commerce
27 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Natus Brain Monitor Breakout Box, Part Number 021911; full-montage standard electroencephalograph

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI/DI 038283NA00153ED, Serial Numbers: 021918100008-M, 021918100014-M, 021918100015-M, 021918100017-M, 021918100019-M, 021918100020-M, 021918100021-M, 021918100025-M, 021918100026-M, 021918100028-M, 021918100032-M, 021918100034-M, 021918100036-M, 021918100038-M, 021918100039-M, 021918100040-M, 021918100041-M, 021918100044-M, 021918100046-M, 021918100047-M, 021918100048-M, 021918100049-M, 021918100050-M, 021918100051-M, 021918100052-M, 021918100054-M, 021918100055-M.

View the original FDA recall notice

Why was this product recalled?

Natus received two complaints of left side labeling show impedance for right side.

What actions did Natus Neurology DBA Excel Tech., Ltd. (XLTEK) take?

Natus issued an Urgent Medical Device Recall notice to its consignees on 06/18/2025 via USPS. The notice explained the issue, risk, and requested the user immediately cease use and quarantine pending return of the unit. A replacement device(s) will be made available to you at no cost, once you provide the completed Customer Reply Form that is listed below and return to FCA@natus.com. Technical Service will be in contact with you to arrange for a provision of replacement device(s).

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology